Table 3 Showing the descriptive data of operative and postoperative parameters

From: Impact of prosthesis oversizing on clinical outcomes of transcatheter aortic valve implantation using a self-expandable Evolut R valve

Operative data

No. = 143

DI (mm)

Median (IQR)

3.1 (2.3–4.85)

 

Range

1–8

DI I (mm/m2)

Median (IQR)

1.69 (1.25–2.43)

 

Range

0.42–4.6

Pre- implantation Dilatation

No

107 (74.8%)

 

Yes

36 (25.2%)

Post implantation-Dilatation

No

118 (82.5%)

 

Yes

25 (17.5%)

DIMS %

Median (IQR)

36.14 (24.62–59.62)

 

Range

10–114.23

ECG post operative

No.

%

Conduction disturbance

No

113

79.0%

 

Yes

30

21.0%

Permanent CHB

No

141

98.6%

 

Yes

2

1.4%

Transient conduction. Disturbance

No

132

92.3%

 

Yes

11

7.7%

New LBBB

No

119

83.2%

 

Yes

24

16.8%

QRS widening

No

114

79.7%

 

Yes

29

20.3%

PR Prolongation

No

137

95.8%

 

Yes

6

4.2%

New onset AF

No

140

97.9%

 

Yes

3

2.1%

Postoperative data

No. = 143

DIDO %

Mean ± SD

16.55 ± 5.93

 

Range

0–30

PVL

No (0)

61 (42.7%)

Trivial (1)

68 (47.6%)

Mild (2)

12 (8.4%)

Moderate ( 3)

2 (1.4%)

Severe (4)

0 (0.0%)

Coronary obstruction or annular rupture

No

143 (100.0%)

  1. AF Atrial fibrillation, CHB Complete heart block, DI Depth of implantation, DIi Depth of implantation indexed, DIMS Depth of implantation to membranous septum ratio, LBBB Left bundle branch block, DIDO Depth of implantation derived oversizing index, PVR Para-valvular leakage