Skip to main content

Table 3 correlation between maximum and minimum ASD diameter measured by 3D echocardiogram and diameters measured by 2D TTE and 2D TEE

From: A novel three-dimensional echocardiographic method for device size selection in patients undergoing ASD trans-catheter closure

 

2D TTE ASD PSSX

2D TTE ASD biatrial

2D TTE ASD sagittal

2D max diameter

TEE max diameter

3D d1

r

0.829

0.790

0.723

0.780

0.895

P

<0.0001

<0.0001

<0.0001

<0.0001

<0.0001

3D d2

r

0.871

0.826

0.730

0.792

0.902

P

<0.0001

<0.0001

<0.0001

<0.0001

<0.0001

  1. r= Pearson correlation coefficient