Skip to main content

Table 2 Comparison and correlations between OBP, ABPM (average 24 h and daytime), and ASqBP measurements, means ± SD

From: The use of automated sequential blood pressure in hypertension clinics compared with office and ambulatory blood pressure measurements

  OBP ABPM (average 24 h) R P value
 SBP (mmHg) 142.7 ± 15.5 134.0 ± 15.4 0.64 > 0.001
 DBP (mmHg) 88.5 ± 12.3 80.2 ± 13.2 0.66 < 0.001
  OBP ABPM (daytime) R P value
 SBP (mmHg) 142.7 ± 15.5 135.8 ± 15.7 0.64 > 0.001
 DBP (mmHg) 88.5 ± 12.3 82.1 ± 13.7 0.66 < 0.001
  OBP ASqBP R P value
 SBP (mmHg) 142.7 ± 15.5 137.0 ± 16.8 0.81 > 0.001
 DBP (mmHg) 88.5 ± 12.3 86.4 ± 13.8 0.83 > 0.001
  ABPM (average24 h) ASqBP R P value
 SBP (mmHg) 134.0 ± 15.4 137.0 ± 16.8 0.74 > 0.001
 DBP (mmHg) 88.5 ± 12.3 86.4 ± 13.8 0.74 > 0.001
  ABPM (daytime) ASqBP R P value
 SBP (mmHg) 135.8 ± 15.7 137.0 ± 16.8 0.74 > 0.001
 DBP (mmHg) 82.1 ± 13.7 86.4 ± 13.8 0.73 > 0.001
  1. ABPM ambulatory blood pressure measurement, ASqBP automated sequential blood pressure, DBP diastolic blood pressure, OBP office blood pressure, SBP systolic blood pressure