Outcome of ductus arteriosus stenting including vertical tubular and convoluted tortuous ducts with emphasis on technical considerations

Abstract

Background

Ductal stenting is the preferred method of securing adequate pulmonary blood flow in patients with duct-dependent pulmonary circulation. The main limitation in most centers is the difficult vertical tubular or convoluted ducts that represent real challenges to interventional pediatric cardiologists. We present our experience in patent ductus arteriosus (PDA) stenting with some technical tips to overcome difficulties, especially in stenting tortuous or long tubular ducts. This study was conducted on all patients with cyanotic congenital heart disease who underwent PDA stenting between January 2011 and December 2018.

Results

We attempted to stent the PDA in 43 patients, with a success rate of 93% (40 patients) and only one procedural mortality. There was also one stent migration that needed to be treated with urgent surgery. Three-fourths of the patients had difficult ductal morphology and origin. One stent was used to cover the PDA in 27 patients (62.8%), two stents were used in 13 (30.2%), and three stents were used in 2 patients (4.6%). In-stent stenosis rate was 12.5% (5 patients) and the development of progressive left pulmonary artery stenosis was seen in two patients (5%). Pulmonary artery growth was adequate in all patients.

Conclusions

PDA stenting is an effective method of palliation for patients with duct-dependent pulmonary circulation. It has low morbidity and mortality rates. Stenting difficult ducts have become more feasible with evolving materials and techniques.

Background

Stenting of the patent ductus arteriosus (PDA) is the primary treatment for patients with duct-dependent pulmonary circulation. It serves as a temporary bridge for later surgical repair [1]. It is considered a better alternative to conventional shunt surgery, owing to the latter’s high morbidity and mortality [2]. Two large studies found that PDA stenting was superior to Blalock–Taussig (BT) shunt placement, with fewer complications, shorter hospital stay, less use of diuretics, and better growth of pulmonary arteries [3, 4]. The outcome of ductal stenting has improved due to great advancements in coronary stent materials and design, and the improvement of guidewires and catheters, which in turn has led to greater success rates. However, the technical challenges of stenting long tubular or convoluted ducts are still a limiting factor in some centers [1]. Therefore, interventional pediatric cardiologists continue to modify their techniques of approaching, crossing, and stenting difficult ducts.

This study aimed to present our experience in patent ductus arteriosus (PDA) stenting with some technical tips to overcome difficulties, especially in stenting tortuous or long tubular ducts. Short- and mid-term outcomes of the procedure were also presented.

Methods

This study included all patients with cyanotic congenital heart disease who underwent cardiac catheterization to stent the PDA between January 2011 and December 2018.

Pre-procedure assessment

A detailed echocardiography was performed with special emphasis on the origin, morphology, and size of the PDA. Since most of our patients were neonates with a single source of pulmonary blood flow (PDA), intravenous prostaglandin administration was adjusted to the lowest possible dose required to maintain oxygen saturation at around 75%. We did not stop the prostaglandin except when the duct was approached, crossed with the wire (and occasionally with the stent), and angiography was performed, to accurately assess the morphology and size of the narrowest diameter of the PDA and to avoid any spasm while crossing.

Procedure

The procedure was usually performed under general anesthesia, particularly in newborns. However, deep conscious sedation was occasionally used in older infants. Using the Seldinger technique, the femoral artery and vein were both accessed with a 4F sheath and a 5F sheath, respectively. Our approach to ductal stenting depended on ductal morphology, particularly, the origin of the ductus from the aorta, and the presence of a ventricular septal defect (VSD). The retrograde approach through the femoral artery was used in most patients, while the antegrade approach through the femoral vein was used in a few patients with tetralogy of Fallot/pulmonary atresia (TOF/PA), especially if the PDA originated from the undersurface of a right aortic arch. In those patients, the PDA was crossed either from the left side of the heart through a patent foramen ovale (PFO) or from the right side through a VSD. Both techniques are risky due to the possibility of either mitral valve injury, or variable degrees of heart block and arrhythmias while going through the VSD, especially in patients with congenitally corrected transposition of the great arteries (CCTGA).

Angiography was usually performed with a Judkins’ right (JR4) or pigtail catheter to determine the morphology, length, and diameter of the PDA, as well as the confluence of the pulmonary arteries (PAs) and the presence of stenosis. Ducts were classified into 4 groups according to their origin: arising from the proximal descending aorta (group A); arising from the undersurface of the proximal arch, i.e., vertical (group B); intermediate origin (group C); and arising from the subclavian or brachiocephalic artery (group D) [1].

In PDAs originating from the aortic arch undersurface, which represented most patients, we cut the pigtail catheter tip to form an inverted J to engage the ductal ampulla. However, we have recently adopted the use of the pigtail catheter without cutting the tip, as PDA spasm is more likely to occur due to ductal intimal injury by the sharp edge. Therefore, the new technique adopted involved placing the 4F Fep Pigtail® Merit Medical over the wire, allowing the leading part of the wire to straighten (open) the pigtail’s rounded distal end. Once at the arch, we withdrew the leading wire segment slowly back into the pigtail catheter and then pulled back the catheter slowly to drop into the PDA orifice with its normal soft rounded tip without causing ductal injury. Lately, the left internal mammary artery (LIMA) 4F catheter, which has a stiffer tip than the pigtail, had also been used to carefully engage the ampulla of the PDA. When the PDA arose from the proximal descending aorta or the arch branches, a JR4 catheter was maneuvered to engage the ampulla.

In the antegrade approach through the VSD to the aortic arch, a 5F Judkins’ right guiding catheter was preferred in accessing the PDA, as it provided good support when placing the stent and confirming its position through angiography using a Y-connector.

Once the catheter had been directed toward the PDA ampulla, multiple coronary wires (without any preference) were used to cross and were deployed into the PAs. Attempts resulting in straightening of the duct and ductal shape change with crossed wires mandated a repeat angiography, and the length of the PDA was re-measured to obtain the final stent length needed.

In the retrograde approach, the coronary stent was manipulated over the wire from the femoral artery access point up to the PDA without using long sheaths; based on fixed and stable anatomical and radiological reference points that have been predetermined before deployment. These points included the vertebral body and the intervertebral disk borders, the tracheal air column, the clavicle, and the tip of the endotracheal tube.

The type of stent used depended upon its availability. The stent needed to cover the entire ductal length from the pulmonary to the aortic ends, adding an extra 2–3 mm as a safety margin to ensure total coverage. The diameter of the stent was selected based on the patient’s weight and PDA diameter. A single stent between 15 and 23 mm was likely to be deployed in most patients. A long stent was likely to cause difficulty in tracking the wire, especially in tortuous or vertical tubular ducts. Therefore, instead of using a long stent, a shorter stent was deployed in the distal ductus followed by a second overlapping stent for the proximal part.

The first stent covered the lower curve of the duct from the pulmonary end into its mid-portion, with the second overlapping up to the subclavian artery to ensure coverage of the entire ductal length. Occasionally, a third stent was needed to cover the PDA origin at the subclavian artery, if left uncovered by the second stent.

Post-procedure

After the procedure, all patients were kept ventilated for at least 6 h and given Enoxaparin 1 mg/kg/dose twice daily for 24 h, in addition to oral aspirin 5 mg/kg/day on the same night. Most of the patients were kept nil per orem (NPO) for 24 h after the procedure, especially low weight neonates, and feeding was gradually introduced as tolerated.

Follow-up

The patients were seen regularly in the outpatient clinic, with the assessment of oxygen saturation and echocardiography to assess stent patency and pulmonary artery anatomy. Cardiac catheterization and angiography were indicated in patients with decreased oxygen saturation. The interventions were dilation, re-stenting of the PDA stent or diagnostic pre-surgical assessment, either for further repair or palliative procedures.

Echocardiography was performed before the procedure for diagnosis, evaluation of PDA size, shape, and origin. It was also done after the procedure for evaluation of ductal stent patency and pulmonary arterial growth during the follow-up period.

Statistical analysis

Statistical analysis was performed using SPSS Version 26 software (IBM SPSS Statistics for Windows, Version 26.0. Armonk, NY: IBM Corp). Data were expressed as median and range (minimum–maximum) for numerical data, number, and percentage for categorical data. Comparison before and after intervention for oxygen saturation was tested using the Wilcoxon test. P value < 0.05 was considered of statistical significance.

Results

PDA stenting was attempted in 43 patients (Table 1). The median age was 18 days (range 1 day to 6 years, with 9 patients after the neonatal period). The median weight was 3 kg (range 2.1–13 kg). Five patients (11.6%) were preterm, and 18 patients (41.9%) were female. The majority had pulmonary atresia with VSD (TOF/PA) (27.9%, 12/43), while pulmonary atresia with the intact interventricular septum (PA/IVS) represented 20.9% (9/43). The remaining patients had single ventricle physiology with pulmonary atresia or severely limited pulmonary blood flow.

Table 1 Patients’ characteristics

The retrograde approach was used in most patients (38/43, 88.4%), while the antegrade approach was used in five patients (11.6%). Nearly one-third of our patients had a right aortic arch (30.2%, 13/43). The median PDA diameter was 2.4 mm (range 0.8–3.9 mm) and the median PDA length was 17.3 mm (range 9–35 mm).

Ducts originating from the proximal descending aorta were found in 14 patients (32.5%), from the subclavian artery in 10 (23.2%), and the undersurface of the aortic arch in 37.2% (16/43). Three of our patients had double PDAs (all are successfully stented). The ducts were tortuous in 53.5% (23/43) and were vertical tubular ducts in 27% (12/43). PDAs originating from the left subclavian artery in the patients with a right aortic arch were noted to be C-shaped and were the longest ducts in our cohort (Fig. 1). These PDAs were usually 30–36 mm in length, and rarely less than 26 mm. For these ducts, using a very long stent would have probably lead to failure of the procedure or increase its risks, as the stent would straighten the C-shaped duct and stretch the PAs and the subclavian artery, increasing the likelihood of injury. Therefore, in such ducts, we used two or three stents (Fig. 1). Table 1 illustrates the demographic and angiographic characteristics of studied patients.

Fig. 1
figure 1

Stenting a long tubular patent ductus arteriosus (PDA) from the subclavian artery. Upper panel images A–D show one stent used to cover the PDA. Lower panel images E–H show the use of 2 stents

Successful PDA stenting was achieved in 40 patients with a success rate of 93%. In one patient, the stent migrated to the branch PAs, while the procedure failed in another with three overlapping stents. Both these patients were sent for surgical palliation. The third patient died before crossing the duct with the stent.

One stent was used to cover the entire length of the duct in 27 patients (62.8%), while two stents were used in 13 patients (30.2%). Only two patients (4.6%) used three stents: one of them had a double PDA (Fig. 2) and the other one had a right aortic arch and a very long C-shaped duct from the left subclavian artery. The median stent diameter was 3.5 mm, and the median stent length was 18 mm (Table 2).

Fig. 2
figure 2

Stenting of double patent ductus arteriosus (PDA) in a patient with tetralogy of Fallot/pulmonary atresia and disconnected pulmonary arteries. The right pulmonary artery is supplied by a PDA from the subclavian artery that was stented with 2 stents (upper panel images, A–C). The left pulmonary artery is supplied by a PDA from the proximal descending aorta that was stented with one stent (lower panel images, D–F)

Table 2 Cardiac catheterization data of studied patients

Bare-metal stents (BMS) were preferred over drug-eluting stents (DES), as there still is not enough evidence regarding the off-label use of DES in such cases. Integrity bare-metal stents (Medtronic; Minneapolis, Minnesota) represented 37.9% of the stents used in our study. Multi-link vision stents (Abbott Vascular; Abbott Park, Illinois) was used in 41.2%, Challenger (Synexmed; Shenzhen, China) in 10.3%, Driver (Medtronic; Minneapolis, Minnesota) in 5.1%, Resolute Onyx (Medtronic; Minneapolis, Minnesota) in 3.4%, and PALMAZ BLUE™ (Cordis, Cardinal Health; Dublin, Ireland) in 1.7% of our patients. The median fluoroscopy time for the procedure was 13 min (range 5–119 min) and the median contrast volume was 10.5 ml.

Immediate outcome

The median arterial oxygen saturation increased from 75 to 85% (P = 0.001).

There was one mortality due to the closure of the PDA. The patient died before the wire was passed or the stent was placed. The procedure was performed, as an emergency, on a critically ill baby with severe acidosis who was transferred from another hospital (Table 2).

Stent migration caused cardiac arrest in one patient who recovered after cardiopulmonary resuscitation. The stent was retrieved with concomitant BT shunt placement. Another patient was extubated accidentally and developed respiratory arrest. After difficult re-intubation and stabilization, the procedure was completed with no in-hospital sequelae. Sudden transient desaturation with no stent obstruction was observed in two patients (4.8%). One patient developed reperfusion injury due to excessive pulmonary over circulation.

Hospital course

The median duration of hospital stay was 5 days (range 1–278 days), with an in-hospital mortality rate of 15% (n = 6). One patient died due to acute stent thrombosis, two patients due to severe pneumonia, and three patients from fulminant sepsis. In-stent stenosis rate was 12.5% (5 patients) as seen in Fig. 3.

Fig. 3
figure 3

In-stent stenosis. This is a patient with tricuspid atresia/pulmonary atresia and double patent ductus arteriosus (PDA) (A), both stented due to severe confluence stenosis (B), presented a few months later with desaturation and in-stent stenosis (C). Re-stenting with 2 more stents (D, E) was done successfully (F)

Development of left pulmonary artery (LPA) stenosis after stenting was detected during cardiac catheterization in two patients, while pre-existing LPA stenosis did not show significant progression until the time of surgical repair in two TOF patients with preserved growth of the pulmonary arteries (Fig. 4). The rest of the patients showed adequate growth of the PAs by echocardiography or cardiac catheterization. Data for every patient were presented in a separate table/supplementary material (Additional file 1).

Fig. 4
figure 4

Left pulmonary artery (LPA) stenosis and growth of pulmonary arteries associated with ductal stenting. A The vertical duct from the undersurface of the aortic arch, B The regular conical duct from the proximal descending aorta. C, D Pre-existing LPA stenosis. On routine follow-up angiography, there was a development of stent stenosis and LPA stenosis (E, F). Follow-up angiography images show adequate growth of pulmonary arteries despite right pulmonary artery jailing (G, H)

Discussion

PDA stenting has emerged as an alternative to surgical shunt palliation. It has the advantages of being less invasive, thereby avoiding sternotomy and cardiopulmonary bypass, and potentially allowing for more symmetrical growth of the pulmonary arteries [5], aside from the shorter hospital stay compared to surgery [4]. However, PDA stenting in patients with duct-dependent pulmonary blood flow is still a Class IIb indication (level of evidence: C) according to the 2011 American Heart Association guidelines, and the presence of branch pulmonary artery stenosis is a Class III indication (level of evidence: C) [6].

Stenting long vertical or convoluted ducts always represents a challenge to interventional cardiologists and is a real limitation of PDA stenting in some centers [1]. Despite this, we were able to successfully palliate 40 patients with duct-dependent pulmonary blood flow (out of 43 attempts) via PDA stenting instead of creating surgical aortopulmonary shunts, irrespective of PDA origins and morphologies. We did not have any exclusion criteria concerning weights, ages, age of gestation (preterm), or anatomical diagnoses, unlike other reports that excluded tortuous, vertical, or very long ducts [7], as well as pulmonary artery stenosis [8].

Our practice is to keep the intravenous prostaglandin running until the duct is crossed with the wire. This technique may decrease the ductal spasm or enlarges the stenotic area while manipulating the wire in crossing trials, increasing the success rate of placing the wire deep in the lungs.

In our cohort, stenting was performed through femoral access without aggressive axillary or carotid arterial cannulation, or cutdown [9, 10], which we've observed to be the usual access point of engaging long and tortuous ducts with multiple curves and constrictions in other centers. Recent generations of softer and more floppy wires, flexible catheters with better torque, and stents with better profile and flexibility played major roles in safe, successful stenting through the femoral route.

Buys et al. [9] reported 80–100% procedural success while avoiding stenting tortuous and vertical ducts. Despite having a large proportion of tortuous and vertical ducts in our cohort (approximately 80%), our success rate was 93%, which is comparable to that of Alwi and other studies [7, 8, 11].

For convoluted tortuous ducts, it is important to straighten the duct (Fig. 5) by crossing the stump of the main pulmonary artery with the wire and then deflecting it back to the branch pulmonary artery, instead of crossing directly from the PDA to the branch PA (Fig. 6). The stents placed in a tortuous duct might change their orientation and position once the wire support is taken, which might result in uncovered ductal tissue constricting and causing desaturation. This can be avoided by straightening the duct with the wire and placing the stent based on the straightened length and adding a few millimeters to be within the aortic lumen, keeping in mind that the stent will cross the curves of the ductus and will not follow the initial tortuosity.

Fig. 5
figure 5

Stenting a convoluted duct. The upper panel images (A–C) show the tortuous duct with the wire taking the same complete circle of the duct. The wire was manipulated to straighten the duct (D). Two stents were used to cover the whole length of the duct (E–F)

Fig. 6
figure 6

Stenting a tortuous duct. (A) Angiography showing a tortuous duct. The wire was pushed to the pulmonary artery stump then reflected to one of the pulmonary branches (B) to straighten the duct and stent it safely (C–D)

When it is difficult to cross the convoluted ducts with one long stent, stenting the distal pulmonary end first with a short stent, then overlapping with a second stent to cover the rest of the duct until the aortic end can be done. As studied and proven by Baho et al. [12], stent protrusion slightly into the pulmonary artery and sometimes the aorta causes no harm compared to leaving a part of the duct unstented.

Ductal length is of paramount importance, as underestimating it might leave an uncovered area that becomes stenotic while overestimating it may result in significant protrusion into the pulmonary artery or aorta.

It is very unlikely that a single stent of 15 mm or less in length will cover the entire PDA length. Most of these patients have underestimated PDA length due to tortuosity. Our single case of stent migration was in a patient stented with a short stent (< 15 mm).

We used more than one stent to adequately cover the PDA length in 15 patients (34.8%). Most of the trials used only one stent to cover the ductal length [7, 9, 11, 13]. Alwi et al. [8] used two stents in only one of 56 patients (1.7%). Two stents were used in a few of our patients because it is sometimes difficult to measure the length accurately with convoluted and angled ducts.

Three stents were used in two patients. One patient had disconnected PAs and a double PDA (Fig. 2); the right pulmonary artery (RPA) was supplied by a stenotic duct from the right subclavian artery, and the LPA was supplied by a regular duct from the aortic arch. Two stents were implanted in the ductus coming from the subclavian artery, and one stent was used for the other ductus. The other patient who required three stents had a right aortic arch with a long C-shaped ductus coming from the left subclavian artery.

Stents typically used for PDA stenting are a flexible and newer generation of coronary artery stents. About 80% of the stents used in our study were bare-metal stents. Drug-eluting stents have been reported to result in less in-stent restenosis than BMS when deployed in the PDA in an animal model [14]. Preliminary data on the use of DES and pharmacokinetics in neonates suggest significantly lower clearance of sirolimus in neonates and peak sirolimus levels being 20 times higher than in older children and adults [15]. Further studies are needed to assess the clinical relevance of this theoretical systemic immunosuppression.

Our reported periprocedural complication rates were low. Major complications included stent migration, cardiac arrest that required resuscitation, and subacute thrombosis. We did not have any case of permanent femoral vessel damage, contrary to a study by Glatz wherein this occurred in 9% of patients [3].

After the procedure, the neonates were kept on mechanical ventilation for 6–24 h to avoid acute reperfusion injury to the lungs. NPO was maintained for 24 h to avoid the possibility of gut ischemia after feeding due to the steal phenomenon from the PDA. Normal feeding was allowed after ensuring an absence of significant pulmonary over circulation.

There were a few in-hospital deaths (6 patients, 13.9%), mostly due to nosocomial sepsis and pneumonia. Most of the trials had a few deaths before hospital discharge, generally unrelated to stent implantation [7, 8, 10, 16,17,18,19,20].

In-stent stenosis has been reported to have a variable rate [8,9,10, 12] ranging from no restenosis at 6 months [21] up to 60%. This exceptionally high rate has been attributed to small stent sizes (3 and 3.5 mm) and incomplete coverage of the ducts [10]. Alwi had a stenosis rate of 16% [8]. In our group of patients, stent stenosis was detected in five patients (12.5%). The patients were successfully managed with catheter-directed re-interventions only. Two patients required repeat stenting for restenosis, and another two required balloon dilatations only. Repeated balloon dilatation or re-stenting has managed to maintain acceptable pulmonary blood flow to palliate inoperable patients for a period close to 4 years.

Pre-existing LPA stenosis is not uncommon, particularly in TOF/PA. Many investigators believe that PDA stenting tends to accelerate pre-existing branch PA stenosis [17, 22]. Alwi had seven patients with pre-existing mild LPA stenosis that progressed after stent implantation [8]. We only had two patients with pre-existing LPA stenosis, and we proceeded with stent implantation, with no progression in the degree of stenosis. Only two of our patients (nearly 5%) developed LPA stenosis, while the remaining patients had adequate growth of their PAs.

Partial or complete jailing of a branch PA was noted occasionally in our patients. It was noted in 22% of patients in one study [23]. However, in the long term, there was no significant difference in branch PA size and symmetry, regardless of whether or not a branch PA was jailed [22].

Limitations of the study

The limitation of our study was largely the small number of cases, and this is explained by the rarity of such anomalies. It is also limited by its retrospective design.

Conclusions

Patent ductus arteriosus stenting is a reliable first choice palliative strategy for patients with duct-dependent pulmonary blood flow. Difficult cases with long, vertical, or tortuous ducts can be approached safely and stented successfully. Using more than one stent is feasible and safe, to overcome difficulties and ensure complete coverage of the duct. Ductal patency can be maintained with adequate arterial oxygen saturation and balanced pulmonary artery growth until the patients can get their definitive repair or palliation.

Availability of data and materials

All data were available at King Abdulaziz University Hospital.

Abbreviations

BMS:

Bare-metal stent

CCTGA:

Congenitally corrected transposition of great arteries

DES:

Drug-eluting stents

IVS:

Intact ventricular septum

LIMA:

Left internal mammary artery

LPA:

Left pulmonary artery

NPO:

Nil per orem

PA:

Pulmonary atresia

PAs:

Pulmonary arteries

PDA:

Patent ductus arteriosus

RPA:

Right pulmonary artery

TOF:

Tetralogy of Fallot

VSD:

Ventricular septal defect

References

  1. Alwi M (2008) Stenting of ductus arteriosus: case selection, technique, and possible complications. Ann Pediatr Card 1(1):38–45

    Article  Google Scholar 

  2. Petrucci O, O’Brien SM, Jacobs ML, Jacobs JP, Manning PB, Eghtesady P (2011) Risk factors for mortality and morbidity after the neonatal Blalock–Taussig shunt procedure. Ann Thorac Surg 92(2):642–651

    Article  Google Scholar 

  3. Glatz AC, Petit CJ, Goldstein BH, Kelleman MS, McCracken CE, McDonnell A et al (2018) A comparison between patent ductus arteriosus stent and modified Blalock–Taussig shunt as palliation for infants with ductal-dependent pulmonary blood flow: insights from the congenital catheterization research collaborative. Circulation 137(6):589–601

    Article  Google Scholar 

  4. Bentham JR, Zava NK, Harrison WJ, Shauq A, Kalantre A, Derrick G et al (2018) Duct stenting versus modified Blalock Taussig shunt in neonates with duct-dependent pulmonary blood flow: associations with clinical outcomes in a multicenter national study. Circulation 137(6):581–588

    Article  Google Scholar 

  5. Santoro G, Gaio G, Palladino MT, Iacono C, Carrozza M, Esposito R et al (2018) Stenting of the arterial duct in newborn with duct-dependent pulmonary circulation. Heart 94(7):925–929

    Article  Google Scholar 

  6. Feltes TF, Bacha E, Beekman RH, Cheatham J, Feinstein JA, Gomes AS et al (2011) Indications for cardiac catheterization and intervention in pediatric cardiac disease. Circulation 123(22):2607–2652

    Article  Google Scholar 

  7. Matter M, Almarsafawey H, Hafez M, Attia G, Abuelkheir MM (2013) Patent ductus arteriosus stenting in complex congenital heart disease: early and midterm results for a single-center experience at children hospital, Mansoura, Egypt. Pediatric Cardiol 34(5):1100–1106

    Article  Google Scholar 

  8. Alwi M, Choo KK, Abdul Latiff H, Kandavello G, Samion H, Mulyadi MD (2004) Initial results, and medium-term follow-up of stent implantation of patent ductus arteriosus in duct-dependent pulmonary circulation. J Am Coll Cardiol 44(2):438–445

    Article  Google Scholar 

  9. Buys DG, Brown SC, Greig C (2013) Stenting the arterial duct: practical aspects and review of outcomes. SA Heart 10(3):514–519

    Google Scholar 

  10. Schneider M, Zartner P, Sidiropoulos A, Konertz W, Hausdorf G (1998) Stent implantation of the arterial duct in newborns with duct-dependent circulation. Eur Heart J 19(9):1401–1409

    Article  CAS  Google Scholar 

  11. Raval A, Thakkar B, Madana T, Oswal N, Garg R, Umalkar R et al (2016) Ductus arteriosus stenting: a promising percutaneous palliation in patients with duct-dependent pulmonary circulation. Rev Port Cardiol 35(11):583–592

    Article  Google Scholar 

  12. Baho H, Deraz S, Abouzeid H, Al-Ata J, Kouatli A (2013) Stent protrusion in palliative congenital heart disease interventions: does it cause any harm? J Invasive Cardiol 25(9):460–463

    PubMed  Google Scholar 

  13. Francis E, Kumar S, Kumar RK (2014) Palliative stenting of patent ductus arteriosus in older children and young adults with congenital cyanotic heart disease. Catheter Cardiovasc Interv 83(7):1109–1115

    Article  Google Scholar 

  14. Lee KJ, Hinek A, Chaturvedi RR, Almeida CL, Honjo O, Koren G, Benson LN (2009) Rapamycin-eluting stents in the arterial duct: experimental observations in the pig model. Circulation 119(15):2078–2085

    Article  CAS  Google Scholar 

  15. Lee KJ, Seto W, Benson L, Chaturvedi RR (2015) Pharmacokinetics of sirolimus-eluting stents implanted in the neonatal arterial duct. Circ Cardiovasc Interv 8(5):e002233

    Article  Google Scholar 

  16. Hussain A, Al-Zharani S, Muhammed AA, Al-Ata J, Galal OM (2008) Midterm outcome of stent dilatation of patent ductus arteriosus in ductal-dependent pulmonary circulation. Congenit Heart Dis 3(4):241–249

    Article  Google Scholar 

  17. Michel-Behnke I, Akintuerk H, Thul J, Bauer J, Hagel KJ, Schranz D (2004) Stent implantation in the ductus arteriosus for pulmonary blood supply in congenital heart disease. Catheter Cardiovasc Interv 61(2):242–252

    Article  Google Scholar 

  18. Gewillig M, Boshoff DE, Dens J, Mertens L, Benson LN (2004) Stenting the neonatal arterial duct in duct-dependent pulmonary circulation: new techniques, better results. J Am Coll Cardiol 43(1):107–112

    Article  Google Scholar 

  19. Vigo CM, Pinedo GM, Mondragon CS, Montes ML, Manrique SA, Bejarano GH et al (2015) Immediate and medium-term outcomes of ductal stenting in neonates and infants. Rev Bras Cardiol Invasiva 23(3):211–215

    Google Scholar 

  20. Santoro G, Capozzi G, Caianiello G, Palladino MT, Marrone C, Farina G et al (2009) Pulmonary artery growth after palliation of congenital heart disease with duct-dependent circulation. J Am Coll Cardiol 54(23):2180–2186

    Article  Google Scholar 

  21. Amoozgar H, Cheriki S, Borzoee M, Ajami G, Soltani M, Ahmadipour M et al (2012) Short-term result of ductus arteriosus stent implantation compared with surgically created shunts. Pediatr Cardiol 33(8):1288–1294

    Article  Google Scholar 

  22. Alwi M (2011) Patent ductus arteriosus stenting problems, complications and technical consideration. Neonatol Today 6(7):1–5

    Google Scholar 

  23. Qureshi AM, Goldstein BH, Glatz AC, Agrawal H, Aggarwal V, Ligon RA et al (2019) Classification scheme for ductal morphology in cyanotic patients with ductal dependent pulmonary blood flow and association with outcomes of patent ductus arteriosus stenting. Catheter Cardiovasc Interv 93(5):933–943

    Article  Google Scholar 

Download references

Acknowledgements

Not applicable.

Funding

No fund was received for this study.

Author information

Authors and Affiliations

Authors

Contributions

SB: Concept and design, JA: Concept and design, NA: Data collection, AA: Revision of the manuscript, ZZ: Revision of manuscript, BA: Data collection, MA: Data collection and drafting of initial manuscript, AE: Data collection, AD: data interpretation, GA: Statistical analysis and critical revision of the manuscript. All authors have read and approved the manuscript.

Corresponding author

Correspondence to Gaser Abdelmohsen.

Ethics declarations

Ethics approval and consent to participate

The study was approved by the ethical committee of King Abdulaziz university hospital. Written consents from the patients ′ legal guardians were taken retrospectively on admission for potential participation in future research and publication. (committee’s reference number: 763-21).

Consent for publication

Not applicable.

Competing interests

The authors declare that they have no competing interests.

Additional information

Publisher's Note

Springer Nature remains neutral with regard to jurisdictional claims in published maps and institutional affiliations.

Supplementary Information

Additional file 1

. Appendix to manuscript.

Rights and permissions

Open Access This article is licensed under a Creative Commons Attribution 4.0 International License, which permits use, sharing, adaptation, distribution and reproduction in any medium or format, as long as you give appropriate credit to the original author(s) and the source, provide a link to the Creative Commons licence, and indicate if changes were made. The images or other third party material in this article are included in the article's Creative Commons licence, unless indicated otherwise in a credit line to the material. If material is not included in the article's Creative Commons licence and your intended use is not permitted by statutory regulation or exceeds the permitted use, you will need to obtain permission directly from the copyright holder. To view a copy of this licence, visit http://creativecommons.org/licenses/by/4.0/.

Reprints and permissions

About this article

Check for updates. Verify currency and authenticity via CrossMark

Cite this article

Bahaidarah, S., Al-Ata, J., Alkhushi, N. et al. Outcome of ductus arteriosus stenting including vertical tubular and convoluted tortuous ducts with emphasis on technical considerations. Egypt Heart J 73, 83 (2021). https://doi.org/10.1186/s43044-021-00210-4

Download citation

  • Received:

  • Accepted:

  • Published:

  • DOI: https://doi.org/10.1186/s43044-021-00210-4

Keywords